Victoreen 07-494
Operators Manual

1-2

1.2 Specifications

Range

Low: 50 kVp to 90 kVp

High: 80 kVp to 150 kVp

Resolution

0.1 kVp

Accuracy

± 3% or 3 kV, whichever is greater (tungsten target x-ray tube
with 4.5 mm Al total filtration)

mAs Requirements

See Figure 1-2

Reproducibility

± 1 kV or 1%, whichever is greater

Filtration Effects

50 kVp - 90 kVp range: less than 3.5% with 5.3 mm aluminum
added

80 kVp – 150 kVp range: less than 3% with 8 mm aluminum
added

Calibration Frequency

One (1) year

Battery

Type: 9 V alkaline, NEDA 1604 (IEC 6F22) or alkaline equivalent

Life: 75 hr. typical

Display

3 ½ digit, 0.5" high LCD with LO BAT indicator;
Below Range kVp: Flashing middle decimal point

Above Range kVp: Flashing left decimal point

Above Range mAs: Flashing right decimal point

Below Range mAs: Flashing right decimal point and 00.0
display

LED Indicators

Auto Reset: New exposure detected, last kVp reading cleared

Range Fault: Range select sliding filter not properly engaged

Controls

Power Switch: ON/OFF

Phase Selection Switch: Single phase/3 phase

Filter Range Selection Switch: 50 to 90 kVp or 80 to 150 kVp

Connectors

BNC connector for waveform output

Dimensions
(H x W x D)

2.5 in. x 8 in. x 6 in.

(6.4 cm x 20.3 cm x 15.2 cm)

Weight

2.4 lbs. (1.1 kg)

Environmental Specifications

Temperature:

Operating: 59 to 104° F (15 to 40° C)

Storage: - 0 to 122° F (-18 to +55° C)

Relative Humidity:

Operating: 0 to 90% non-condensing

Storage: 0 to 90% non-condensing