background image

Victoreen 07-494
Operators Manual

5-4

7. Adjust R6 until the display reads 85 kVp.

8. Set the x-ray machine to 140 kVp, 50 mA, and 0.3 seconds.

9. Make an exposure.

NOTE


If the C/D decimal point is flashing, indicating that
the intensity of the exposure is too high, decrease
the mA or increase the FDD and repeat the
exposure.

10. Adjust R3 until the display reads 140 kVp.

11. Repeat Steps 5 through 10 until the readings are accurate.