Examples, Us standard metric – Floscan FloNET Standard Flow Model User Manual
Page 33
01/04/11
2000-085-00C
. SIMULATOR MODE:
I
Rotating the Green Switch to position number 5, (US Gallons) or position letter D, (Liters) places the FloNET module into Simulator
Mode. While in Simulator Mode the module generates a simulated fuel flow signal, (PGN 127489 Engine Parameters Dynamic) with
small fluctuations onto the display’s upper line. The signal is also transmitted
onto the NMEA 2000 backbone and can be used to
heck display compatibility and NMEA 2000 cable connections.
c
he lower line shows a simulated fuel total, (GAL TOT or LIT TOT). The current Instance number is displayed to the right of TOT.
T
EXAMPLES
:
US STANDARD
METRIC
FloScan Instrument Company, Inc.
Tel:
(206)
524-6625
Fax:
(206)
523-4961
3016 NE Blakeley Street, Seattle, WA 98105
Email:
Http://www.floscan.com
: POSITION D:
POSITION 5
46.0 GPH (SIM)
1.4 GAL TOT 0
175.0 LPH (SIM)
5.3 LIT TOT 0
II. DIAGNOSTIC TEST:
1.
Start the engine(s) and allow them to “warm-up” for 5 to 10 mi
es.
nut
2.
With engine(s) running, place the FloNET Module into Diagnostic Test mode. To do so rotate the Green Switch to Position 1,
(for US Gallons models) or Position 9, (for Metric models). This places the instrument into Diagnostic Test mode. The display’s
upper line shows raw, (un-calibrated) fuel flow through the Forward Sensor. The display’s lower line shows raw, (un-calibrated)
fuel flow through the Return Sensor.
OTE: Calibration settings do not affect diagnostic mode readings.
N
EXAMPLES
:
US STANDARD
METRIC
: POSITION 9:
POSITION 1
F 497.6 GPH
R 518.3 GPH
F 1883 LPH
R 1962 LPH
3.
d 1800 RPM. Circle
Using the chart below, record Forward and Return sensor readings at IDLE, 1000 RPM, 1500 RPM an
S or F to indicate steady or fluctuating readings. If fluctuating, show fluctuation range, (High & Low).
Forward
Return
S F
IDLE
S F
S F
1000
S F
S F
1500
S F
S F
1800
S F
4. Rotate the Green Switch through Positions 2 & 3 for US Gallon models or A & B for Metric Models. Record the various readings
in the chart below the example boxes.
EXAMPLES
:
POSITION 2:
POSITION A:
F 75 F FUEL TMP
R 90 F FUEL TMP
F 24 C FUEL TEMP
R 32 C FUEL TEMP
- N20D-201-2K FloNET N20D-BOS-2K FloNET N20RBBOS-2K FloNET N20D0-2012K FloNET N20D-231-2K FloNET N20D0-BOS2K FloNET N20D0-2312K FloNET N20RB201-2K FloNET N2TD-235-2K FloNET FloNET Hi Capacity Flow Model N2TD-6DB-2K FloNET N20D-3CB-2K FloNET N2TD-6DC-2K FloNET N2TD-6DD-2K FloNET N2TRB6DB-2K FloNET N20RB3CB-2K FloNET N2TD-6CB-2K FloNET N2TRB6CB-2K FloNET N2TD0-6DD2K FloNET N2TD-6ED-2K FloNET N20D-3BB-2K FloNET N20D0-3EE2K FloNET N20D-3DC-2K FloNET N20D0-3BB2K FloNET N20D0-3CB2K FloNET N2TD0-6CB2K FloNET N2TD0-6ED2K FloNET N2TD-6GG-2K FloNET N2TD-6FD-2K FloNET N20D0-3ED2K FloNET N2TD-6CC-2K FloNET N2TD-6FE-2K FloNET N20D0-3DB2K FloNET N2TD0-6FE2K FloNET N2TD0-6BB2K FloNET N2TD-6BB-2K FloNET N2TD-6EE-2K FloNET