Electro-Chemical Devices (ECD) HYDRA Ammonium User Manual
Page 22

Page 14
HYDRA NH
4
+
-N
3.2.7 Status Menu
The Status menu displays the actual uncompensated millivolt value from the sensor, very handy
information for troubleshooting. The day and time, hours: minutes: seconds, are also displayed.
Screen Displayed
Button Pressed
Ch1 Status
◄
CALIBRATE ► (enter screen)
Fri 09:11:28
MENU SELECT ▼
Input 364.2 mV
CALIBRATE ▼ (return to Home Screen)
3.2.8 Configuration and Trim Menu
The Configure/Trim menu provides access to the menus used to assign parameters, define functions and
trim input and output signals. This group of menus should rarely be needed as the instrument was
configured at the factory before testing and shipment. The basic structure is as follows. See the “System
22 Configuration Manual” for details about these menus, available at www.ecdi.com.
Screen Displayed
Button Pressed
Configure/Trim
◄
CALIBRATE ► (enter screen)
Passwords
MENU SELECT ▼(allows menus to be password protected)
Default Display
MENU SELECT ▼(defines Home Screen)
4-20 assign
MENU SELECT ▼(Assigns parameter/function to mA output)
Manual Mode
MENU SELECT ▼(Allows manual control of mA outputs)
Relay assign/Test
MENU SELECT ▼(Assigns triggering parameter to a relay)
PID Assign
MENU SELECT ▼(Assigns PID to a PV)
PWM Assign
MENU SELECT ▼(Assigns Pulse Width Modulation to a PV)
Clock and timers
MENU SELECT ▼(Sets Clock function, assigns Timers)
Logic Gates
MENU SELECT ▼(assigns AND/OR gates)
Ion Species
MENU SELECT ▼(Sets Ion Measured)
General
MENU SELECT ▼( Not Used)
°C/°F & Temp Cal
MENU SELECT ▼( allows temperature trim/cal, choose °C/°F)
4-20 Trim/Test
MENU SELECT ▼(allows mA trim)
Millivolt Trim
CALIBRATE ▼ (return to Home Screen)
3.3 OUTPUTS
The Standard C-22 HYDRA Analyzer has two 4-20 mA outputs. The default configuration has “mA 1”
configured for NH4-N, 4-20 mA = 00.1 ppb to 50 ppm and “mA 2” is configured for pH, 4-20 mA = 2-12
pH. To reconfigure the outputs see Section 3.2.6, Set-Up Menu, above.