8.0 ENGINEERING DOCUMENTATION
The information and technical data disclosed in this document may be used and disseminated only for
the purposes and to the extent specifically authorized in writing by Electro-Chemical Devices. Electro-
Chemical Devices reserves the right to change published specifications and designs without prior notice.
8.1 SPECIFICATIONS
Sensors and Flow Train
TRO (Total Residual Oxidant Sensor)
Combined sensor, Platinum electrode, KCl/AgCl reference electrode with porous Teflon junction, 3 K-
ohm temperature sensor and signal conditioner
PHS 10 (pH sensor)
Combined sensor, pH glass electrode, KCl/AgCl reference electrode with porous Teflon junction, 3 K-
ohm temperature sensor and signal conditioner
Measurement Range
TRO
±1500 mV
pH
2 - 12 pH
Temperature
0° to 60° C (32°- 122°F)
Pressure
50 psig, max. (3.4 bar, max.)
Wetted Materials
PVC, Poly Propylene, PTFE, Viton, Glass, Platinum
Process Connections
Input
Schedule 80 PVC, ¾” slip fitting
Drain
Schedule 80 PVC, ¾” slip fitting
Performance Characteristics
Measurement Range:
0.00 to 2.00 ppm Chlorine
Accuracy:
Dependent on the accuracy of the DPD test and pH value
Minimum Detection Limit:
0.014 ppm (MDL)
Limit of Quantification:
0.044 ppm (LOQ)
Response Time:
Chlorine < 2 minutes (T90)
Temp. < 2 minutes (T90)
pH < 30 seconds (T90)
Drift
< 1.5% month