Electro-Chemical Devices (ECD) DCA-23 User Manual

Page 17

background image

DCA-23

Page 11

3.3 CALIBRATION MENUS

3.3.1 pH Calibration

Screen Displayed

Button Pressed

1 pH 7.00 pH

Cal .0 mV

CALIBRATE ►(enters calibration line, zero pt. Cal 1)

Sensor should be in a 7.0 pH solution.

1 pH 7.00 pH

Cal 0.2 mV

MENU SELECT ▼ (accepts the Calibration when the mV

value is stable)

1 pH 7.00 pH

Cal 0.2 mV

MENU SELECT ▼ (move down to Cal line)

(Entering the Cal line resets value to 0.0 mV)

1 pH 7.00 pH

Cal 0.2 mV

MENU SELECT ▼(move to Cal 2, Slope)

Remove and rinse sensor with distilled water

2 pH 10.00 pH

Cal 59.1 mV/pH

CALIBRATE ► (enter Cal Line)

Place Sensor in a pH 10.00 solution

2 pH 10.00 pH

Cal 60.2 mV/pH

Set the pH value using the CALIBRATE ◄ or ► to move

cursor. Use the ▲or▼ to adjust the integer value.

2 pH 10.00 pH

Cal 60.2 mV/ppm

MENU SELECT ▼ (accept Calibration)

2 pH 10.00 pH

Cal 60.2 mV/ppm

MENU SELECT ▲ (return to Home Screen)

3.3.2 TRO Calibration
Screen Displayed

Button Pressed

1 TRO 180.0 mV

Cal 173.4 mVa

Run DPD test on Sample, Determine TRO value from

Calibration Curve using DPD Value and pH

1 TRO 180.0 mV

Cal 173.4 mVa

press ◄ CALIBRATE ► to initiate Calibration.

1 TRO 526.0 mV

Cal 519.0 mVa

Enter the TRO Value using the CALIBRATE ◄ or ► to move

cursor. Use the ▲or▼ to adjust the integer value.

1 TRO 526.0 mV

Cal 519.0 mVa

MENU SELECT ▲ (accept Calibration)

2 TRO .00 mV

Cal 1.000 mV/TRO

MENU SELECT ▲

Verify slope 1.000 mV/TRO

2 TRO .00 mV

Cal 1.000 mV/TRO

CALIBRATE ► (enter Cal Line)

Resets value to factory Default: 1.000 mV/TRO

2 TRO .00 mV

Cal 1.000 mV/TRO

MENU SELECT ▼ (return to Home Screen)

3.3.3 Temperature Trim

Screen Displayed

Button Pressed

Trim °C .0

Unit 12.8°C

press ◄ CALIBRATE ► to initiate Calibration.

Trim °C 2.4

Unit 15.2°C

Enter the Temp Value using the CALIBRATE ◄ or ► to

move cursor. Use the ▲or▼ to adjust the integer value.