DCA-23 Dechlorination Analyzer

3.1 KEYS

The blinking cursor indicates the active point where menus can be selected or numerical values

adjusted. There are two sets of keys on the DCA-23 analyzer, the MENU SELECT keys and the CALIBRATE

keys.

The MENU SELECT keys are used to move the cursor vertically changing the displayed menu. These keys

are also used to Save/Accept the calibration data and exit the calibration menu.

The CALIBRATE keys are used to enter menus, change numerical values and move the cursor.

To enter a calibration menu or parameter adjustment line, simultaneously press both of the

Horizontal CALIBRATE keys, ◄CALIBRATE►

Pressing either of the Horizontal CALIBRATE keys separately will move the cursor horizontally to the

point under the digit to be adjusted. The Vertical CALIBRATE keys are used to adjust numeric values.

Pressing the upper key will increase the value and pressing the lower key will decrease the value.

3.2 MENUS

There are two sets of menus on the Model DCA-23. The Channel 1 menus (PV1) are the TRO menus, the

Channel 2 menus (PV2) are the pH menus. Pressing the MENU SELECT Down key will access the pH

channel, pressing the MENU SELECT Up key will access the TRO channel. See the guide below for the

general menu structure.

3.2.1 TRO Menus

Screen Displayed

Button Pressed

Cl ppm .23

23.6% 12.8°C

Home Screen (view only)

Press MENU SELECT ▲

to move to TRO Screen

TRO 526.9 mV

12.8°C

TRO Screen (view only)

MENU SELECT ▲

Input 520.5 mVa

0 mVa 6.4 mV

TRO Raw Input Screen (view only)

MENU SELECT ▲

1 TRO 180.0 mV

Cal 173.5 mVa

TRO Calibration 1 Screen

MENU SELECT ▲

2 TRO 0.0 mV

Cal 1.000 mV/TRO

TRO Calibration 2 Screen

MENU SELECT ▲

4 mA .00 ppm

20 mA 1.00 ppm

Chlorine 4-20 mA Adjustment screen

MENU SELECT ▲

ON .00 ppm
OFF .00 ppm

Chlorine Alarm Relay Adjustment screen

MENU SELECT ▲