beautypg.com

0 troubleshooting – Electro-Chemical Devices (ECD) DCA-23 User Manual

Page 23

background image

DCA-23

Page 17

6.0 TROUBLESHOOTING

The DCA-23 was evaluated and calibrated at the factory before shipment. Upon initial start up the
system should require minimal to no adjustments.
Verify the system has sample flow and the pH and the temperature sensors are reading correctly. These

parameters affect the measurement and must be reading properly. If these conditions are met and

problems still exist use the Troubleshooting Table to find a remedy.

Troubleshooting Guide

Symptom

Possible Cause

Remedy


Displayed value is Higher
than DPD test value.


pH sensor reading to high


TRO sensor reading to high

Check mV value in pH 7
buffer, if > -60 mV replace the
electrode cartridge. If < -60mV
perform a one pt. Cal
Check the TRO value vs. the
tested DPD chlorine value on
Calibration Curve. Correct if
necessary.


Displayed value is Lower than
DPD test value


pH sensor reading to low


TRO sensor reading to low

Check pH of grab sample and
standardize the pH reading

Check the TRO sensor in a
ORP standard solution, if
more than 60 mV low then
replace the electrode, other-
wise recalibrate to DPD test.


Zero Chlorine Reading in a
solution with measureable
chlorine.


TRO sensor electrically
shorted

Check TRO sensor in a ORP
standard solution
Electrode failure or sealing o-
ring failure. Remove electrode
cartridge from sensor and
look for water inside the
sensor, replace electrode and
sensor if water inside.


Unstable Chlorine Reading


Mostly Air bubbles in line

Oscillating pressure > 5 psi

Coating on pH or TRO sensor


Remove air from line.

Dampen pulses with
accumulator before flow cell
Oil or grease on pH or TRO will
cause noisy readings, clean
electrode tips with detergent