Beurer GL 50 (mg/dL) User Manual
Page 33

Beurer GL50 mg/dL
33
Note
The blood glucose monitor is made of precision components. The accuracy of the measurements
and service life of the device depend on its careful handling:
• Protect the device from impacts and do not drop it.
• Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or
water, marked temperature fluctuations, direct sunlight and extreme cold.
• Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile
telephones.
8 w
hat
if
there
are
problems
?
Display messages on batteries and blood glucose measurement
No.
Cause
Solution
LP
Batteries empty.
Replace all batteries.
Ht
Temperature of the measuring envi-
ronment, blood glucose monitor or
test strip above the permitted range.
Repeat the test using a new test strip as soon
as the measuring environment, blood glucose
monitor and test strips have reached room tem-
perature (+20°C to +26°C).
Lt
Temperature of the measuring envi-
ronment, blood glucose monitor or
test strip below the permitted range.
Repeat the test using a new test strip as soon
as the measuring environment, blood glucose
monitor and test strips have reached room tem-
perature (+20°C to +26°C).
Err
Used or contaminated test strip in-
serted.
• Insert an unused test strip that has not ex-
pired.
• Repeat the blood glucose measurement.
Err 001
System error.
Remove batteries, reinsert batteries. Contact
Customer Services if the problem persists.
Err 002
Insufficient blood on the test strip.
Repeat the measurement using a new test
strip.
Err 005
System error.
Remove batteries, reinsert batteries. Contact
Customer Services if the problem persists.
Unknown error messages.
Remove batteries, reinsert batteries. Contact
Customer Services if the problem persists.