Beurer GL 50 (mg/dL) User Manual
Page 23

Beurer GL50 mg/dL
23
3
Stick protective disc on needle
Place the retained protective disc flat on a hard surface. Stick the tip
of the needle into the protective disc (image 1) so the needle is cov-
ered (image 2).
Image 1 Image 2
4
Remove and dispose of lancet needle
Press the trigger again so that you can grip the shaft. Carefully remove
the lancet needle from the lancing device and dispose of the lancet in
a puncture-proof container.
Carefully dispose of all blood samples and materials that you have come
into contact with. This prevents injuries and the infection of others.
5
Replace cap
Place the cap back on the device. Turn the clip back over the cap.
5.6 Evaluating measured blood glucose values
Your blood glucose monitor can process values between 20 and 630 mg/dL. The “Lo” warning is
displayed for measured results below 20 mg/dL. The “ ” warning is displayed for measured results
above 630 mg/dL.
Warning
• If you suspect that the blood glucose results are incorrect, first repeat the test and, if applicable,
perform a functional test using control solution. Seek medical advice if dubious results persist.
• Seek medical attention immediately if your symptoms do not correspond to your measured blood
glucose values and you have followed all instructions for the Beurer GL50 blood glucose monitor.
• Do not ignore symptoms of too high/low blood glucose levels. Always seek medical attention!