12. Technical specifications
Model no.
BM 85
Measurement me-
thod
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
pulse 40 –180 beats /minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /
diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 180 mm x W 99 mm x H 40 mm
Weight
Approximately 317 g (without cuff)
Cuff size
22 to 36 cm
Permissible operating
conditions
+ 10 °C to + 40 °C, ≤ 90 % relative humi-
dity (non-condensing)
Permissible storage
conditions
-20 °C to + 55 °C, ≤ 90 % relative humi-
dity, 800 –1050 hPa ambient pressure
Power supply
DC 5 V
600mA lithium ion battery
3.7 V/400mAh
Battery life
For approx. 50 measurements, depen-
ding on levels of blood pressure and
pump pressure
Accessories
Cuff, cuff holder, instructions for use,
USB mains part, USB cable, storage
pouch
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
Data transfer via
Bluetooth® wireless
technology
The blood pressure monitor uses Blue-
tooth® Smart (low energy),
2.4 GHz frequency band,
Compatible with Bluetooth 4.0 smart-
phones/tablets
List of supported smartphones/tablets