Cleaning and care, Technical details – Beurer BC 32 User Manual
Page 18
18
• The cuff is fastened too tightly or loosely:
E3;
• Errors occur during the measurement:
E4;
• The pump pressure is higher than 300 mmHg:
E5;
• The batteries are almost empty
E6.
In such cases, repeat the measurement. Ensure that you do not
move or speak. If necessary, reinsert or replace the batteries.
8. Cleaning and Care
• Clean your blood pressure monitor carefully using a slightly
damp cloth only.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.
9. Technical details
Model no.
BC 32
Type
BC32/1
Measurement
method
Oscillometric, non-invasive blood
pressure measurement on the wrist
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 30 – 200 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 72 mm x W 69 mm x H 29.2 mm
Weight
Approx. 101 g (without batteries)
Cuff size
135 to 195 mm
Permissible operating
conditions
+5 °C to +40 °C, 15-93 % relative air
humidity (non-condensing)
Permissible storage
conditions
-25 °C to +70 °C, 10-93 % relative air
humidity, 700 –1060 hPa ambient pres-
sure
Power supply
2 x 1.5 V
AAA batteries
Battery life
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA batter-
ies, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notification
to allow for updates.