beautypg.com

Cleaning and care, Technical details – Beurer BC 32 User Manual

Page 18

background image

18

• The cuff is fastened too tightly or loosely:

E3;

• Errors occur during the measurement:

E4;

• The pump pressure is higher than 300 mmHg:

E5;

• The batteries are almost empty

E6.

In such cases, repeat the measurement. Ensure that you do not
move or speak. If necessary, reinsert or replace the batteries.

8. Cleaning and Care

• Clean your blood pressure monitor carefully using a slightly

damp cloth only.

• Do not use any detergents or solvents.

• Never hold the instrument under water as otherwise liquid

can penetrate and damage the instrument.

• Never place any heavy objects on the instrument.

9. Technical details

Model no.

BC 32

Type

BC32/1

Measurement

method

Oscillometric, non-invasive blood

pressure measurement on the wrist

Measurement range

Cuff pressure 0 – 300 mmHg,

systolic 50 – 250 mmHg,

diastolic 30 – 200 mmHg,

Pulse 40 –180 beats/minute

Display accuracy

Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement

inaccuracy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Memory

2 x 60 memory spaces

Dimensions

L 72 mm x W 69 mm x H 29.2 mm

Weight

Approx. 101 g (without batteries)

Cuff size

135 to 195 mm

Permissible operating

conditions

+5 °C to +40 °C, 15-93 % relative air

humidity (non-condensing)

Permissible storage

conditions

-25 °C to +70 °C, 10-93 % relative air

humidity, 700 –1060 hPa ambient pres-

sure

Power supply

2 x 1.5 V

AAA batteries

Battery life

For approx. 250 measurements,

depending on the blood pressure level

and/or pump pressure

Accessories

Instruction for use, 2 x 1.5 V AAA batter-

ies, storage box

Classification

Internal supply, IPX0, no AP or APG,

continuous operation, type BF applied

part

Technical information is subject to change without notification

to allow for updates.