Beurer BC 32 User Manual
Page 12
12
Operating
5°C
40°C
RH 15-93%
Permissible operating temperature and
humidity
Keep dry
SN
Serial number
0483
The CE labelling certifies that the product
complies with the essential requirements of
Directive 93/42/EEC on medical products.
Advice on use
• In order to ensure comparable values, always measure your
blood pressure at the same time of day.
• Before every measurement, relax for about five minutes.
• If you want to perform several measurements on the same
person, wait five minutes between each measurement.
• Do not take a measurement within 30 minutes after eating,
drinking, smoking or exercising.
• Repeat the measurement if you are unsure of the measured
value.
• The measurements taken by you are for your information
only – they are not a substitute for a medical examination!
Discuss the measurements with your doctor, and never base
any medical decisions on them (e.g. medicines and their
administration)!
• Do not use the blood pressure monitor on newborns or pati-
ents with preeclampsia. We recommend consulting a doctor
before using the blood pressure monitor during pregnancy.
• In the case of restricted circulation on the arm as a result of
chronic or acute vascular diseases (including vascular con-
striction), the accuracy of the wrist measurement is limited.
In this case you should avoid using an upper arm blood
pressure monitor.
• Cardiovascular diseases may lead to incorrect measurements
or have a detrimental effect on measurement accuracy. The
same also applies to very low blood pressure, diabetes, circu-
latory disorders and arrhythmias as well as chills or shaking.
• The blood pressure monitor must not be used in connection
with a high-frequency surgical unit.
• Only use the unit on people who have the specified wrist
measurement for the device.
• Please note that when inflating, the functions of the limb in
question may be impaired.
• During the blood pressure measurement, blood circulation
must not be stopped for an unnecessarily long time. If the
device malfunctions, remove the cuff from the arm.
• Do not allow sustained pressure in the cuff or frequent meas-
urements. The resulting restriction of the blood flow may
cause injury.
• Ensure that the cuff is not placed on an arm in which the
arteries or veins are undergoing medical treatment, e.g. intra-
vascular access or therapy, or an arteriovenous (AV) shunt.
• Do not use the cuff on people who have undergone a mas-
tectomy.
• Do not place the cuff over wounds as this may cause further
injury.
• The blood pressure monitor can only be operated with bat-
teries.