Programming with external voltage sources -19 – AMETEK DCS-E 1KW Series User Manual
Page 41

Sorensen DCS Series 1kW and 1.2kW Supplies
Installation and Operation
M362500-01 Rev H
2-19
2.11.1 Programming With External Voltage Sources
The output voltage can be programmed using either a 0-5 Vdc or 0-10 Vdc external voltage
source. To program the output voltage with a 0-5 Vdc source, set switch SW1-3 open (default
factory setting) and remove the jumpers connecting pins 8 to 9 and 20 to 21 on connector J3.
Connect the external source between pins 9 (positive) and 12 (return). Varying the external
voltage from 0-5V will cause the output to vary from 0-100% of rated output.
S
W
1
-1
S
W
1
-2
S
W
1
-3
S
W
1
-4
S
W
1
-5
S
W
1
-6
S
W
1
-7
S
W
1
-8
CLOSED
CONNECTOR J3
0-5V
1312
25
9
1
14
SWITCH SW1
OPEN
Switch SW1
SW1-1 Off (Open)
SW1-5 Off (Open)
SW1-2 Off (Open)
SW1-6 Off (Open)
SW1-3 Off (Open)
SW1-7 Off (Open)
SW1-4 On (Closed)
SW1-8 On (Closed)
Switch SW1 and Connector J3 Configuration
for 0-5V Programming of the Output Voltage
(J3 sense line, OVP and current control jumpers shown set for local operation)
For programming with a 0-10 Vdc source, close switch SW1-3 and replace the 0-5V source with
a 0-10V source.
S
W
1
-1
S
W
1
-2
S
W
1
-3
S
W
1
-4
S
W
1
-5
S
W
1
-6
S
W
1
-7
S
W
1
-8
CLOSED
CONNECTOR J3
0-10V
1312
25
9
1
14
SWITCH SW1
OPEN
Switch SW1
SW1-1 Off (Open)
SW1-5 Off (Open)
SW1-2 Off (Open)
SW1-6 Off (Open)
SW1-3 On (Closed)
SW1-7 Off (Open)
SW1-4 On (Closed)
SW1-8 On (Closed)
Switch SW1 and Connector J3 Configuration
for 0-10V Programming of the Output Voltage
(J3 sense line, OVP and current control jumpers shown set for local operation)