Quickcheck mode – Etymotic ER•200DW8 Personal Noise Dosimeter with Data Logging User Manual

Page 13

background image

13

13

QuickCheck Mode

(Estimated Dose per Hour) Test Time: 2 minutes

Power ON

· Press and release the power button

three times

.

- The 12.5% (green) LED illuminates on the first press.
- The 25% (green) LED illuminates on the second press
- The 50% (yellow) LED illuminates on the third press.

· The LEDs sweep twice from bottom to top.
· The bottom LED flashes green throughout the

measurement, indicating QuickCheck mode.

· The solid LED indicates the predicted dose per hour.


After 2 minutes the LEDs sweep twice from bottom to top,
indicating the measurement is complete.

· The solid LED (predicted dose per hour) displays

for 7 seconds while the bottom LED flashes green
(indicating QuickCheck). The dosimeter then shuts off.

· Result is stored in the device memory and is displayed

when the device is turned on again in Status mode.
Stored data can also be downloaded to a PC.

Power OFF

To power off prior to QuickCheck’s automatic shutdown
you must wait at least 15 seconds into the measurement.

· Press and hold the power button for 2 seconds. The LEDs

will sweep twice, bottom to top.

· The estimated dose per hour (solid LED) is displayed

for 7 seconds while the bottom LED flashes green
(indicating QuickCheck). The dosimeter shuts off.

NOTE: If the dosimeter is shut off prior to its automatic shutdown, it
will still calculate the predicted hourly dose based on the sound level
and duration of the measurement.