beautypg.com

Interpreting results, Normal mode, Led dose% risk factor – Etymotic ER•200DW8 Personal Noise Dosimeter with Data Logging User Manual

Page 12: Based on niosh criteria)

background image

12

Interpreting Results

Normal Mode

(Based on NIOSH Criteria)

LED

Dose%

Risk Factor

Hearing

Protection

Needed

green 12.5%

(slow flash)

< 12.5%

No risk of hearing

loss

No

green 12.5%

(regular flash)

12.5 - 24%

No risk of hearing

loss

No

green 25%

25 - 49%

No risk of hearing

loss

No

yellow 50%

50 - 99%

1/2 or more of

daily dose

Yes

(conservatively)

red 100%

100 - 199%

Exposure limit

exceeded

Yes

red 2x

200 - 399%

> 2x allowable

daily dose

Yes

red 4x

400 - 799%

> 4x allowable

daily dose

Yes

red 8, 16, 32x

(regular flash)

8x

> 8x allowable

daily dose

Yes

red 8, 16, 32x

(double flash)

16x

> 16x allowable

daily dose

Yes

red 8, 16, 32x

(fast flash)

32x

> 32x allowable

daily dose

Yes