8 error messages, Technical information 7.1 pump model specification – Casella CEL Apex personal sampling pumps User Manual
Page 32

32
13
Specify Start and End dates and times on the Program Times dialog.
14.
Enable the selected Time (
Ö) so it will be used by the program.
15.
Select and enable further start and end times, then click OK to confirm the selection.
16.
To send the complete package of four programs to the Apex Pro, place the unit in range of the 
infrared transducer, and click the Send to Pump button. 
Please note: It is not possible to send an individual program to the pump. 
The four current programs will be transferred to the pump, while a Pump Programming 
Status dialog gives progress information. 
The PC takes this opportunity to synchronise the clock in the Apex unit to the same time and date 
as the PC. 
17.
Once the program has been sent to the Apex unit, the unit can be removed from the transceiver.
6.8
Error Messages
The following error messages may be displayed by Pump Manager. 
“ERROR: Unable to get status record from the pump.” 
“ERROR: Unable to download record %i.” 
“ERROR: Pump changed during download.” 
“ERROR: Unable to verify the pump’s serial number.” 
“ERROR: Unable to clear the pump’s memory.” 
“ERROR: No pump found to program.” 
“ERROR: Cannot set pump’s time.” 
“ERROR: Cannot set pump’s date.” 
“ERROR: Could not send program to the pump.” 
“ERROR: Could not get the pump’s serial number.” 
“ERROR: The pump has changed during programming.” 
“ERROR: Could not find a pump.” 
“ERROR: Could not open communications.” 
“Invalid serial number” 
“Pump shut down but restarted following a temporary inlet blockage.” 
“Pump terminated the event due to a fatal inlet blockage or flow control problem.” 
“Pump event terminated due to a discharged battery condition.” 
 
7.
TECHNICAL INFORMATION
7.1
Pump Model Specification
Non I.S. versions
I.S. versions
Flow range 
 
 
0.8 to 5 L/min, 
5 to 850 ml/min with low flow 
adaptor, 
0.8 to 4 L/min 
5 to 850 ml/min 
with low flow 
adaptor, 
Flow control accuracy 
 
<±5% for selected flow, 
±3% for calibrated point, 
<±5% for 
selected flow 
±3% for 
calibrated point, 
Flow performance 
(Flow, max vacuum pressure) 
1.0 L/min @ 80 cm H
2
O
2.0 L/min @ 80 cm H
2
O
3.5 L/min @ 32 cm H
2
O
4.0 L/min @ 18 cm H
2
O
5.0 L/min open flow
1.0 L/min @ 80 
cm H
2
O
2.0 L/min @ 55 
cm H
2
O
3.5 L/min @ 16 
cm H
2
O
4.0 L/min open 
flow 
Battery voltage and capacity( Apex Lite utilizes 4 x AA dry cells.)
4.8 V NiMH / 2.7 Ah
4.8 V NiMH / 
1.7 Ah 
Inlet pulsation ratio
< 10% using Dewell Higgins Cyclone @ 2 L/min 
approx. 7.5 cm H
2
O.
Satisfies EN1232 , NIOSH 0600
Displayed values 
 
Real flow rate, volume sampled, elapsed sample 
time, temperature, operating mode, program 
details 
Service interval
Typically 2500 hrs
Operating temperature
5
o
C to 45
o
C
5
o
C to 40
o
C
Storage temperature
-10
o
to + 50
o
C
Charging technique 
 
 
Drop-in Intelligent fast charger 
employing dT/dt termination with 
safety time out. 
Standby trickle charge mode to 
keep pump ready for use. 
 Drop-in 
Intelligent fast 
charger 
employing 
dV/dt 
termination 