Introduction, Presentation – ABUS Technologies Log Box-DA Data Logger User Manual
Page 4

ABUS TECHNOLOGIES INC.
4
Log Box-DA
1. INTRODUCTION
LogBox-DA is an electronic data logger with two input channels, one analog
and one digital. Input data
is stored in the logger internal memory, for later download
to a PC for visualization and analysis in the form of tables or graphs. The data is easily
exported to spreadsheets.
The LogChart-II is the software used to configure the logger, download and
visualize data. The logger configuration allows defining the logger operation mode,
including programming the start/stop time of data acquisition. Other parameters such
as signal input type, logging interval, multiplication factor, scale range, etc, are easily
selected through the LogChart-II software.
The LogBox-DA also provides a signal for commanding an external power
supply (battery) of a device connected to the logger. This feature allows that external
devices, such as a transmitter, be powered only during the measurement sample time,
thus extending the autonomy of these external batteries.
2. PRESENTATION
2.1 Technical Parameters
Inputs:
Channel 1:
NPN, PNP, Pulse or Dry Contact
Level 0:
0.0 up to 1.0 Volts
Level 1:
2.0 up to 24.0 Volts
Max. Frequency:
4000 Hz for NPN, PNP and Pulse
20 Hz for dry-contact (internal De-bounce)
Input Impedance:
NPN, PNP and Pulse: 105 k
Dry contact:
Capacitance of 100 nF
Maximum counting:
65,535 per period.
Channel 2:
0-20mA, 4-20mA, 0-50mV and 0-10Vdc
Input Impedance:
0-20mA and 4-20mA: 120
0-50mV: >>10M
0-10Vdc: >120k
Memory capacity:
32000 (32k) or 64000 (64k)
Interval between readings:
Minimum: 10 seconds, Maximum: 18 hours
External battery switch time:
0 to 10 s
Power:
3.6V lithium battery (1/2 AA)
Estimated autonomy:
Over 2000 days (one daily download)
Working temperature:
From –40°C to 70°C.
Protection:
IP65, IP67 and IP68 models (see lateral label on product)
Material:
ABC + PC case, Polycarbonate film;
Dimensions:
60x70x35mm