ABUS Technologies Log Box-RHT with LCD Data Logger User Manual
Page 14

ABUS TECHNOLOGIES INC.
14
Log Box-RHT
Average:
The value logged will be the mean of ten
consecutive measures taken within a predefined
interval.
The
minimum
interval
between
measurements is 10 seconds.
Maximum:
The value logged will be the maximum value
found within ten consecutive measurements taken
within a predefined interval. The minimum interval
between measurements is 10 seconds.
Minimum:
The value logged will be the minimum value found
within ten consecutive measurements taken within
a predefined interval. The minimum interval
between measurements is 10 seconds.
Offset:
Makes possible to correct the value logged.
Unit:
Defines the unit of the value measured: ºC or ºF
for channel 1 (temperature) and only % for
channel 2 (relative humidity).
Alarm:
Available only in the temperature channel.
Defines limit values that, when exceeded,
characterize an alarm condition. Alarm events are
informed to the user through a flashing mode of
the Alarm Indicator.
Input:
Parameter available only in channel 2 (humidity).
With this parameter it is possible to disable the
humidity measurement, saving all memory
capacity for temperature measurement.
After filling all the fields select OK and settings are sent to the logger.
Palmtop User
To set up the communication between the handheld device and the logger(s), run LogChart
Palm-OS, press the Search Device button from the Recorded Data screen and align the Infrared Port
of the PDA to the logger(s) communication window. If more than one loggers are detected, the Devices
Found screen is exhibited.