ABUS TECHNOLOGIES INC.

12

TagTemp

By date/time: Logging is stopped on user-defined date and time

Interval: Defines the interval between readings: hour, minutes and seconds. When the logging mode is
set to the instantaneous the minimum interval is 1 second. For maximum and minimum values, the
minimum interval is 10 seconds in both models.

Misc: Shows the Logging Mode parameter that defines how measured value will be logged. Options
are:

Instantaneous: The value logged will be exactly the value measured at each interval.
Measurement is taken at the end of the defined interval. The minimum interval between
measurements is 1 second.

Minimum: The value logged will be the minimum value found within ten consecutive
measurements taken within an interval. The minimum interval between measurements is 10
second.

Maximum: The value logged will be the maximum value found within ten consecutive
measurements taken within an interval. The minimum interval between measurements is 10
second.

Alarms: Enables an alarm that is triggered according to user-defined parameters. Values shown in the
temperature unit selected.

PDA Alarm Settings screen

Clocks: Provides access to Logger and PDA clocks. When a new configuration is sent to the logger,
clocks are updated.

Offset: allows fine offset adjustment on the measured value. The Reset button clears changes made
and the logged values are the measured values again.

Offset adjust screen

After configuring clocks in the Settings screen, click Apply to send this configuration to the Logger,
returning to the Monitoring screen.

Sending a new configuration implies the exclusion of all data collected present in the logger memory.