beautypg.com

Nexen TA100 964400 User Manual

Page 9

background image

5

FORM NO. L-20292-C-0700

2.

Connect a voltmeter to TP7 (COM) and TP1 (See
Figure 9).

3.

Rotate S1 clockwise to VDC, increasing the
reading at TP1 (Example: Maximum Tension = 100
Lbs. with a 50 Lb. weight, adjust the output for 5.0
VDC) (See Figure 9).

FIGURE 10

FIGURE 9

S1

S2

TP7

TP2

TP1

FIGURE 11

Jumpers

J3 and J4

S1

TP7

TP2

TP1

SPAN ADJUSTMENT (WITH TWO MB TENSION SENSORS)

1.

Thread a rope or narrow web over the Sensor Roll in the
normal path. Be sure the rope or web is at the center of
the Sensor roll; then, hang a known weight (within the
range of the system) on one end of the rope or web
(Example: Maximum Tension = 100 Lbs. hang a 50 Lb.
weight) (See Figure 10).

2.

Connect a voltmeter to TP7 (COM) and TP1 (See
Figure 11).

3.

Rotate S1 clockwise to VDC, increasing the
reading at TP1 (Example: Maximum Tension = 100
Lbs. with a 50 Lb. weight, adjust the output for 2.5
VDC) (See Figure 11).

4.

Connect a voltmeter to TP7 (COM) and TP2 (See
Figure 11).

5.

Rotate S2 counterclockwise until the voltmeter reads

VDC at TP2 (Example: Maximum Tension =

100 Lbs. with a 50 Lb. weight, adjust the output for
2.5VDC) (See Figure 11).

S2

NOTE

R31 and R32 are factory adjustments only. R31 is a
fine adjustment for the 0-10 VDC. R32 is a fine
adjustment for the 4-20mA.

NOTE

R31 and R32 are factory adjustments only. R31 is a
fine adjustment for the 0-10 VDC. R32 is a fine
adjustment for the 4-20mA.

R31

R32

R31

R32

Jumpers

J3 and J4

This manual is related to the following products: