Detailed Specifications
77
Input Characteristics
Function
Overload
Protection
[1]
Input Impedance
Common Mode
Rejection Ratio
(1 k
Ω
unbalance)
Normal Mode Rejection
L
1000 V
10 M
Ω
<100 pF >120 dB at dc, 50 Hz or 60 Hz
>60 dB at 50 Hz or 60 Hz
F
mV
1000 V
[2]
10 M
Ω
<100 pF >120 dB at dc, 50 Hz or 60 Hz
>60 dB at 50 Hz or 60 Hz
K
1000 V
10 M
Ω
<100 pF
(ac-coupled)
>60 dB, dc to 60 Hz
L
1000 V
3.2 k
Ω
<100 pF
(ac-coupled)
Not specified
Not specified
Full Scale Voltage
Typical Short Circuit Current
Function
Overload
Protection
[1]
Open Circuit
Test Voltage
To 500 k
Ω
>5 or 50 nS
500
Ω
5
k
Ω
50
k
Ω
500
k
Ω
5
M
Ω
50
M
Ω
e
1000 V
[2]
5 V dc
550 mV
<5 V
1 mA
100
μ
A 10
μ
A
1
μ
A 0.3
μ
A 0.3
μ
A
50e 1000
V
[2]
20 V decreasing
to 2.5 V
500 mV
10 mA
G
1000 V
[2]
5 V dc
3.1 V dc
1 mA
[1]
Input is limited to the product of a V rms sinewave times frequency of 2 x 10
7
V-Hz.
[2]
For circuits <0.5 A short circuit. 660V for high energy circuits.