74
Frequency Counter Specifications
Function Range Resolution
Accuracy
99.999 Hz
0.001 Hz
0.02 % + 5
999.99 Hz
0.01 Hz
0.005 % + 5
9.9999 kHz
0.0001 Hz
0.005 % + 5
99.999 kHz
0.001 Hz
0.005 % + 5
[1]
Frequency
(0.5 Hz to 999.99 kHz, pulse
width >0.5
μ
s)
999.99 kHz
0.01 Hz
0.005 % + 5
Duty Cycle
[2][3]
1.00 % to 99.00 %
0.01 %
0.2 % per kHz + 0.1 %
0.1000 ms
0.0001 ms
0.002 ms + 3 counts
1.000 ms
0.001 ms
0.002 ms + 3 counts
10.00 ms
0.01 ms
0.002 ms + 3 counts
Pulse Width
[2][3]
1999.9 ms
0.1 ms
0.002 ms + 3 counts
[1]
For 64.000 kHz to 67.000 kHz, accuracy =
±
5 Hz, at -20 to +55
°
C with R.H. 0 % to 90 % (0
°
C to 37
°
C), 0 % to 65 % (37
°
C to 45
°
C), 0 % to 45 %
(45
°
C to 55
°
C).
[2]
For rise times <1
μ
s. Signals centered around trigger levels.
[3]
0.5 to 200 kHz, pulse width >2
μ
s. Pulse width range is determined by the frequency of the signal.