Ac current specifications – Fluke 289 User Manual
Page 79

True-rms Digital Multimeters
Detailed Specifications
69
AC Current Specifications
Accuracy
Function Range Resolution
20 to 45 Hz
45 to 1 kHz
1 to 20 kHz
20 to 100 kHz
[4]
500
μ
A 0.01
μ
A
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
AC
μ
A
[3]
5000
μ
A 0.1
μ
A
1 % + 5
0.6 % + 5
0.6 % + 10
5 % + 40
50 mA
0.001 mA
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
AC mA
[3]
400 mA
0.01 mA
1 % + 5
0.6 % + 5
1.5 % + 10
5 % + 40
5 A
0.0001 A
1.5 % + 20
0.8 % + 20
3 % + 40
[4]
Not
Spec’d
AC A
[2]
10 A
[1]
0.001 A
1.5 % + 5
0.8 % + 5
3 % + 10
[4]
Not
Spec’d
[1]
10 A range (10 % to 100 % of range).
[2]
20 A for 30 seconds on, 10 minutes off. >10 A not specified.
[3]
400 mA continuous; 550 mA for 2 minutes on, 1 minute off.
[4]
Verified by design and type tests.
See Detailed Specifications introduction for additional information.