34
Light
2. To unlock the control panel, press and hold the “System Lock” button for three (3)
seconds. Ensure the System Lock light extinguishes.
Pulsate Mode
This control unit is equipped with the Pulsate Mode. When selected, the control unit
will continuously readjust the system pressure (proportional to the Comfort Level
setting). Change in pressure is gradual to ensure maximum patient comfort.
To activate, press either “High” or “Low” key (fig. 38) on the control panel. The
Pulsate light will illuminate.
The Pulsate Mode adjusts as follows:
Low–0% cycle every 5 minutes
High–0% cycle under one minute
To deactivate the Pulsate Mode, press the Pulsate
OFF button (Fig. 39).
Note: You may change the rate of alternating pressure at any time while the
Pulsate Mode is in operation.
.
.
3.
Section 3
HybridSelect
Figure 40
Select
Air,
Select
Air Max,
HybridSelect
Models
Flashing
“Green” Light
(Power to
Control Unit)
CPR/Emergency Deflate
Important: In case of a CPR Emergency, turn power OFF by pressing the
CPR button on the keypad.
Figure 39
Note: To assist in deflating the mattress, disconnect the hose at the
mattress or control unit.