ROHO SelectAir MA mattress User Manual
Page 20

18
HealFloat feature (HybridSelect only)
Refer to the ROHO HIGH PROFILE cushionʼs operating manual for specific cleaning
instructions.
Control Unit and Hose (between patients)
1. Unplug the control unit from the wall outlet before proceeding.
2. Wipe down entire control unit and all hoses with a disinfectant solution. MOXI
recommends using LpHse
®
diluted 1:256.
3. Allow solution to remain on system for 10 minutes.
4. Wipe down entire control unit and dry with disposable cleaning cloth or sponge.
5. Only use a mild detergent (dishwashing) and warm water to clean the filter. Rinse
the filter with cool water and thoroughly air dry the filter before replacing.
All Control Units
DANGER:
Disconnect control unit from main power supply before cleaning.
Failure to do so could result in electric shock.
Note:
•Do not immerse the control unit in liquids.
•Do not heat or steam autoclave the control unit.
Note:
Do not saturate the surface of the control unit with liquids. Wipe
off any dust. If necessary, clean the housing exterior with a
disinfectant solution or mild detergent and a damp cloth. Wipe
dry.
DANGER:
Make sure the control unit is completely dry before attempting to
reconnect or operate it.
Section 8
Cleaning and Disinfecting