ROHO SelectAir MA mattress User Manual
Page 27

Figure
Light
Section
Select
Air Max
To unlock the control panel, press and hold the “System Lock” button for three
(3) seconds. Ensure the System Lock light extinguishes.
Auto Comfort
This control unit is equipped with the Auto Comfort mode. When selected, the control
unit will proceed through one or several (depending on the patient's body type) cycles
and will then calculate the approximate proper setting for the patient. If the setting
selected by the control unit is not acceptable, then a second run of the Auto Comfort
cycle may be necessary. If manual setup is desired see Manual Setup, Section 6.
. Turn the control unit on and allow the mattress to fully inflate before making any
adjustments or placing the patient on the mattress.
. Place the patient on the mattress, centered and in their habitual resting position,
i.e. on their back, on their side, etc.
Note: The bed rails should be locked in their highest position before proceeding.
3. Deactivate the Firm Mode before starting Auto Comfort.
4. Press the Auto Comfort “Start” button (fig. 21).
Note: The patient must remain still while the Auto Comfort light is flashing.
Note: Do not press any other keys until the Auto Comfort light is extinguished.
5. Inspect the patient to ensure they have sunk into the mattress about 40-50 % and
not touching the frame at any point. If they have not, adjust the Comfort Level
using the Comfort Level “Softer” and “Firmer” buttons until the proper results
are achieved.
Pulsate Mode
This control unit is equipped with the Pulsate Mode. When selected, the control unit
will continuously readjust the system pressure (proportional to the Comfort Level
setting). Change in pressure is gradual to ensure maximum patient comfort.
. To activate, press either the “High”, “Medium” or “Low” key (fig. 22) on the control
panel. The Pulsate light will come on.
.
Figure
Light
5