34 Installation and Commissioning of the Unit
Operating Manual Freeze Dryer ALPHA 1-4 LDplus / ALPHA 2-4 LDplus
The accessory components are connected to the 
ALPHA 1-4 / 2-4 LD
plus
freeze-dryer according to the connection
diagram. Please observe the following:
Centring rings and clamping flanges with wing nuts are used as 
connecting elements (small-flange connection according to ISO 
28403 or DIN 2861, see the following instructions). 
 
Instructions for the connection with centring rings and 
clamping flanges 
Small flange connections which are not mounted properly between 
the aggregates or to connect hoses are often the reason for serious 
vacuum problems. 
Loosen the connection and place the centring ring (with sealing ring 
inside) in a centred manner between the flange connections.. Seal 
the connection with the clamping ring by tightening the wing nut. 
Ensure that the centring ring neither slips out of place nor gets 
jammed. 
Small flange and centring ring
Small flange with centring ring 
and small flange 
Attaching the clamping ring