12 General Information
Operating Manual Freeze Dryer ALPHA 1-4 LDplus / ALPHA 2-4 LDplus
1.4. Scope of delivery
The scope of delivery includes:
1 tube of high-vacuum grease
1 litre vacuum pump oil (only in case a pump is
delivered)
0.5 m drain hose (silicone 9 x 12 mm)
1 operating manual and further detailed technical
documentation
The scope of delivery does not include:
Commissioning of the unit (inside Germany) can be
performed upon request and will be invoiced at cost.
Installation of the exhaust pipe of the vacuum pump (not
necessary when using an exhaust filter).
1.5. Standards and Regulations
Please refer to the enclosed EU Declaration of Conformity.
1.6. Safety Instructions
1.6.1. CAUTION! Disconnect Mains Plug!
As current-carrying parts are accessible inside the unit
the mains plug must be disconnected before the side panels are
opened or before the control panel is removed.
For maintenance work the unit must be switched off with the mains
switch.
1.6.2. CAUTION! Solvents!
Acidic products or products with a high solvent
concentration cannot be dried without special protective measures
and devices such as e.g. a cooling trap for protection of the vacuum
pump (if necessary check with our service department). Besides,
the unit may be damaged by corrosion.
Special caution is necessary when using azides because a
dangerous explosive develops in combination with copper or non-
ferrous metals! It is absolutely essential to consult our service
department!