Chapter 7: Troubleshooting
Product Service 1-800-522-7658
52
PROBLEM
CAUSE
CORRECTIVE ACTION
Cabinet blower
turns on but lights
don’t work
Lamps not installed
correctly
Inspect lamp installation.
Lamp is defective
Replace lamp.
Keypad
disconnected
or defective
Run keypad diagnostics and check
connections.
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective
lamp
ballasts
Replace lamp ballasts.
Keypad
disconnected
or defective
Run keypad diagnostics and check
connections.
HEPA filter loading
The gauge reading will steadily
decrease as the cabinet is used.
Blockage of the
return air slots or
grille
Check all return air slots and grilles to
ensure that they are not blocked or
restricted.
Blockage of the
exhaust outlet
Ensure that the exhaust outlet is not
blocked or restricted.
Airflow Alert goes
off and/or there is a
slight decrease in
filter life remaining
gauge
Blockage or
restriction under the
work surface
Ensure that the towel catch and plenum
beneath the work surface are
unobstructed.