Refer to the following table if your Bedding Disposal
Station fails to operate properly. If the suggested
corrective actions do not solve your problem, contact
Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Station impellers
and lights won’t
turn on
Unit not plugged into
outlet
Plug the PuriCare into appropriate
electrical service.
Circuit
breakers
tripped
Reset circuit breakers.
Motorized impellers
won’t turn on but
lights work
Impeller wiring is
disconnected
Inspect impeller wiring
Impeller switch is
defective
Replace impeller switch
Motorized impeller is
defective
Replace motorized impeller
Capacitor is defective Replace capacitor
Cabinet impeller
turns on but lights
don’t work
Lamp not installed
correctly
Inspect lamp installation.
Lamp wiring is
disconnected
Inspect lamp wiring