Speed control adjustment – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual
Page 35
![background image](/manuals/738722/35/background.png)
31
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
Fluorescent Lamp Removal:
1. Remove the front dress panel as described above.
2. Locate the control panel on the front of the cabinet,
which has the filter indicator lights and switches.
Remove the screws located on each lower end of
the panel. Swing the panel upward for access to the
fluorescent lamp, as shown in Figure 6-3.
3. Remove the fluorescent lamp by rotating the lamp
and pulling it straight out of its sockets.
4. Install the new lamp by reversing the removal
procedure.
Speed Control Adjustment
The speed control should only be adjusted by a
qualified certifier.
1. Remove the front dress panel as described
above.
2. Adjust the speed control as required by turning
the screw counterclockwise to increase blower
speed, or clockwise, to decrease the blower
speed, as shown in Figure 6-4.
Figure 6-3