Filter indicator system adjustment – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual
Page 36

32
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
3. Replace the front dress panel and reestablish the
inflow air velocity.
Filter Indicator System
Adjustment
The Filter Indicator System was
adjusted at the factory and should not
require recalibration until the HEPA
filter is replaced.
To adjust system, do the following:
1. Ensure that the prefilter is new and the station is
operating within its airflow tolerance.
2. Remove the two screws that secure the front
dress panel, as shown in Figure 6-3, and remove
the panel.
3. Locate the adjustment screw access plug, on the
left side of the enclosure frame, above the
HEPA filter. Remove the plug. Replace the
dress panel.
4. Using cardboard, or a similar material, block off
approximately ½ of the postfilter. This will
simulate an increased load on the system.
Figure 6-4
!