Storage – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual
Page 43

39
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
4. Replace the capacitor with a new one of equal
voltage and capacity.
5. Reassemble the unit by reversing the
disassembly steps. Recertify the enclosure.
Storage
If the Bedding Disposal Station is to be left unused for
more than one month, the unit should be prepared for
storage.
1. Surface disinfect or contaminate the cabinet as
required.
2. Unplug the unit.
3. Cover and seal the work area access opening
and the exhaust outlet with plastic sheeting.
4. Ensure that the cabinet will not be moved or
disturbed while in storage.
The cabinet should not be stored in excessive humidity
or temperature extremes. If the cabinet is moved during
storage it must be recertified before use.