Hepa filter leak test procedure – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual
Page 39

35
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
HEPA Filter Leak Test
Procedure
To ensure dirty room air does not affect the HEPA
Filter Leak Test and that the control panel and sash
gaskets are sealing, take the following steps:
1. Turn on the enclosure and set the average inflow
velocity as indicated in inflow velocity test
procedure. Remove the postfilter
2. Tape closed all openings full length across top,
front and sides of the front dress panel as shown
in Figure 6-7.
3. Turn the photometer on and calibrate it per the
manufacturer’s instructions. The upstream
concentration of DOP will be approximately
24 ug/l. Use these values to establish the
upstream value of 100%.
Figure 6-6
This manual is related to the following products:
See also other documents in the category Labconco Air equipment:
- XPert Weigh Boxes 5220300 (55 pages)
- XPert Balance Enclosures 3930200 (59 pages)
- TXE Tall XPerEnclosure 3892000 (48 pages)
- XPert Filtered Balance Systems 3950200 (81 pages)
- Bulk Powder Enclosures 3963500 (48 pages)
- Biological Safety Cabinets 32239 Series (77 pages)
- Biological Safety Cabinets 30231 Series (61 pages)
- Biological Safety Cabinets 30441 Series (71 pages)
- Biological Safety Cabinets 34309 Series (85 pages)
- Biological Safety Cabinets 34410 Series (85 pages)
- Biological Safety Cabinets 34410 Series (67 pages)
- Biological Safety Cabinets 34300 Series (69 pages)
- Total Exhaust Biological Safety Cabinets 32348 Series (76 pages)
- Total Exhaust Biological Safety Cabinets 30348 Series (64 pages)
- Type C1 Biosafety Cabinet Models 30441 (155 pages)
- Type A1 Biosafety Cabinet Models 30132 (172 pages)
- Type A2 Biosafety Cabinets Models 34300xx (197 pages)
- HEPA Filtered Enclosure Models 3980200 (77 pages)
- HEPA Filtered Enclosure Models 3980200 (78 pages)
- CAptureTMPortable Fuming System 3420000 (29 pages)
- Protecto Downdraft Powder Station 3410000 (25 pages)
- Protecto Series Evidence Drying Cabinets 3390000 (33 pages)
- Protecto Series Evidence Drying Cabinets 3400000 (43 pages)
- Hooded Combination (Macro) Kjeldahl Systems 21176 Series (27 pages)
- Hooded Combination (Macro) Kjeldahl Systems 21179 Series (29 pages)
- Open Combination (Macro) Kjeldahl Systems 21232 Series (27 pages)
- Open Combination (Macro) Kjeldahl Systems 2123Series (25 pages)
- Two and Six-Place (Macro) Kjeldahl Distillation Systems 2128501 (21 pages)
- Two and Six-Place (Macro) Kjeldahl Distillation Systems 2128401 (23 pages)
- Crude Fiber Digestion Apparatus 30001 (16 pages)
- Micro Kjeldahl Digestors 6030000 (17 pages)
- Rapid Digestor Fume Removal System 2350020 (18 pages)
- Rapid Digestors 23012 (30 pages)
- Rapid Distillation Apparatus 65000 (24 pages)
- RapidStill II 65200 (39 pages)
- Protector Controlled Atmosphere Glove Box (80 pages)
- Protector Controlled Atmosphere Glove Boxes 5060000 (53 pages)
- Precise Controlled Atmosphere and Basic Glove Boxes 5220000 (63 pages)
- Precise HEPA-Filtered Glove Boxes 5220200 (54 pages)
- Protector Filtered Glove Box (68 pages)
- AtmosPure Re-Gen Gas Purifiers 5121810 (59 pages)
- HEPA Filter (2 pages)
- Protector Multi-Hazard Glove Box 5065000 (31 pages)
- PuriCare Procedure Stations 31142 Series (62 pages)