Appendix d: oxygen & moisture removal rates – Labconco AtmosPure Re-Gen Gas Purifiers 5121845 User Manual
Page 56

Appendix D: Oxygen & Moisture Removal Rates
Product Service 1-800-522-7658
52
Precise CA (polyethylene) Oxygen Removal with AtmosPure Re-Gen Gas Purifier
7.2
980.2
101.0
30.5
15.8
5.6
5.3
5.1
5.0
4.9
1
10
100
1000
0
50
100
150
200
250
300
350
400
450
Time (Minutes)
O
xy
gen (
PP
M)
This chart shows oxygen level decrease from 980.2 ppm to 4.9 ppm after 400
minutes. Two separate calibrated oxygen monitors confirmed results.
*
*
Results were attained at Labconco Corporation under controlled conditions. Actual results