Appendix d: oxygen & moisture removal rates – Labconco AtmosPure Re-Gen Gas Purifiers 5121845 User Manual
Page 51

Appendix D: Oxygen & Moisture Removal Rates
Product Service 1-800-522-7658
47
Protector CA (Stainless Steel) Moisture Removal with AtmosPure Re-Gen Gas Purifier
1.0
1.8
2.4
3.5
8.5
15.9
11253, 40% RH
1560
87.5
167.5, 220 cycles stop and
AtmosPure circulation blower
starts and isolation valves are
opened.
445
696
1
10
100
1000
10000
100000
1
10
100
1000
10000
Time (Minutes)
Moist
ur
e (PP
M)
This chart shows moisture level decrease from 11,253 ppm (40% relative
humidity) to 1.0 ppm after 8965 minutes. Two separate calibrated moisture
monitors confirmed results.
*
*
Results were attained at Labconco Corporation under controlled conditions. Actual results
may vary.