Review this chapter on maintenance for the following:
1. Routine maintenance.
2. Initial certification.
3. Recertification.
4. Solenoid valve replacement.
5. Catalyst replacement.
Routine Maintenance Schedule
Weekly
Monitor and record oxygen and moisture levels inside the glove box
with accessory monitors (see Chapter 7).
1-3 Months
Regenerate the catalysts every 1-3 months depending on your
application, glove box liner material, and frequency of use.
Initial Certification
Each Gas Purifier has been leak tested at the factory with helium gas at
positive 5" w.g. pressure; no leaks greater than 1 x 10
-6
cc/sec are allowable.
The Gas Purifier and glove box should be certified by your Safety Officer
before use, with one of the two tests listed below under Recertification.