Labconco AtmosPure Re-Gen Gas Purifiers 5121845 User Manual
Page 15
Chapter 3: Getting Started
Product Service 1-800-522-7658
11
Figure 3-1
Rear View of AtmosPure Re-Gen Gas Purifier
C.A. COM Port (6-pin) on back
of Gas Purifier for shared
vacuum pump operation with
glove boxes configured with
Auto Pressure Control feature.
Vacuum Pump Outlet for
glove boxes without auto
pressure control. Used for a
dedicated vacuum pump.
Auto Mode
Moisture/Oxygen
(9-pin) connector for
connection to oxygen
or moisture monitor.
Only used for auto
blower control.
3/8" compression
Exhaust Vent used
during regeneration.
3/8" compression
4% Hydrogen/Inert
Gas Mixture Inlet
3/8" compression
Flow Direction Out
Vent
Solenoid
Valve
Flow Control Valve
to set the 4%
Hydrogen/Inert Gas
Mixture between
10-25 CFH.
Vent Flow Direction
Vacuum Pump
Solenoid Valve
3/4" Inlet/Outlet
Compression
Fittings for Inert
Gas Circulation
from the Glove
Box.