This Appendix contains technical information about all the Gas Purifiers
including electrical specifications and environmental operating conditions.
10 Amps, 100-115V or 5 Amps, 230V, 50/60 Hz AtmosPure Re-Gen
Gas Purifier.
Environmental Conditions
Indoor use only.
Maximum altitude: 12,000 feet (3,657 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the
AC input line that may be as high as 1500V for 100-115V models and
2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally
only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664.