Labconco Protector Controlled Atmosphere Glove Box User Manual
Page 31
Chapter 4: Performance Features and Safety Precautions
Product Service 1-800-522-7658
27
• Performance tested to maintain oxygen levels as low as 0.2 ppm and
moisture levels as low as 1 ppm.
• Oxygen permeation <0.16 ppm/min per ISO 10648-2. Exceeds Class
1 conditions at <1.67 ppm/min.
• Illumination exceeds 80 foot candles.
• Doublewide models include 4 glove ports with main chamber volume
of 34 cu. ft.
3. Basic Manual Control Models include:
• Main chamber pressure analog gauge with readings from -5 to +5
inches of water gauge.
• Transfer chamber pressure analog gauge with readings from 0 to -29.5
inches of mercury.
• Switch to control exterior outlet for vacuum pump connection.
4. Auto Pressure Controller Model include:
• Automatic control of the main chamber pressure (-5 to +5 inches of
water gauge), main chamber evacuation and filling cycles (up to 499
cycles), and transfer chamber evacuation and filling cycles (up to 199
cycles).
• Regulation of vacuum/pressure between -5 and +5 inches of water
gauge in the main chamber and between 0 and -29.5 inches of
mercury in the transfer chamber.
• Mutable, audible alarm that sounds when pressure/vacuum level is
outside set limits.
• LCD display of main chamber and transfer chamber pressure in
inches of water, inches of mercury and mBar.
• Three-position foot pedal (increase gas/fill, off, evacuation/decrease
gas).
• C.A. COM (6-Pin) port for shared vacuum pump use with the optional
AtmosPure™ Re-Gen Air Purifier.
All models feature:
5. Large viewing area, 36" w x 27.6" high, provides maximum visibility.
Removable viewing window is 3/8" thick, laminated safety glass. Double
wide models have two viewing areas.
6. Pair of neoprene gloves, 0.015" thick, 30" long, size 9 ¾. Doublewide
models include two pair. See Replacement Parts for all glove types and
sizes.
7. Space-saving inner and outer transfer chamber doors that pivot upward,
counterbalanced and equipped with quick-latches.
8. Four manual valves – two on the main chamber and two on the transfer
chamber – for evacuating, filling and manual control. With 3/8"
compression fittings and 3/8" hose barbs.
9. Factory-installed, rear-mounted, low profile stainless steel pressure relief
bubbler at ±6 inches of water gauge.
10. Two rear-mounted inlet and outlet 3/4" OD compression fittings for easy
connection to optional AtmosPure Re-Gen Air Purifier or other accessory
drying train.