Labconco Gel Dryer 4330150 User Manual
Page 20

TROUBLESHOOTING
20
PROBLEM
CAUSES
CORRECTIVE
ACTION
Heater and vacuum pump
Gel Dryer fuse
Replace fuse with the
controls don’t work
blown
proper
replacement
Circuit breaker
Reset circuit breaker
tripped
Heater control works but
Vacuum pump
Turn on vacuum
vacuum pump doesn’t
switch in “off”
pump switch
Vacuum pump
Replace fuse with the
fuse blown
proper replacement
Vacuum seal doesn’t
Vacuum leak
Check connection
form, or forms slowly
Insufficient vacuum
Vacuum pump should
pump capacity
have a capacity of
50
liters/minute
or
more
Support screen bent
Ensure that screen
or damaged
lays flat
Silicone rubber
Moisten the surface of
sheet and platen dry
the rubber sheet and
platen
with
water
before
applying
vacuum.
Rubber sheet/platen
Inspect the surfaces
surface damaged
and replace if
necessary
Gel cracks/shatters
Insufficient drying
Increase the drying
time
time
Insufficient vacuum
The vacuum source
should
have
a
minimum
free
air
displacement
of
50
liters/minute
and
develop
a
minimum
vacuum of 25 in. Hg.