Ppendix, Iltered, Nclosure – Labconco Purifier Filtered PCR Enclosures 3970425 User Manual
Page 54: Pecifications

Product Service 1-800-522-7658
50
A
PPENDIX
C
F
ILTERED
PCR
E
NCLOSURE
S
PECIFICATIONS
This Appendix contains technical information about all Purifier Filtered PCR
Enclosures including electrical specifications and environmental operating conditions.
Electrical Specifications
5 Amps, 115V, 60 Hz
3 Amps, 230V, 50/60 Hz
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a temporary
conductivity caused by condensation must be expected, in accordance with
IEC 664.
- Purifier Filtered PCR Enclosures 3970422 Purifier Filtered PCR Enclosures 3970405 Purifier Filtered PCR Enclosures 3970402 Purifier Filtered PCR Enclosures 3970325 Purifier Filtered PCR Enclosures 3970322 Purifier Filtered PCR Enclosures 3970305 Purifier Filtered PCR Enclosures 3970302 Purifier Filtered PCR Enclosures 3970225 Purifier Filtered PCR Enclosures 3970222 Purifier Filtered PCR Enclosures 3970205 Purifier Filtered PCR Enclosures 3970202