Hapter, Roubleshooting and, Ervicer – Labconco Purifier Filtered PCR Enclosures 3970425 User Manual
Page 44: Perating

Product Service 1-800-522-7658
40
C
HAPTER
8
T
ROUBLESHOOTING AND
S
ERVICER
O
PERATING
L
OG
Refer to the following table if your Filtered PCR Enclosure fails to operate
properly. If the suggested corrective actions do not solve your problem, contact
Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Class 100 (ISO Class
5) air conditions are
not present inside the
enclosure.
Improper user
techniques for the
enclosure.
See “Certifying the Enclosure” Chapter 3
and “Safety Precautions” Chapter 4
sections in the manual. (Ref. Appendix D).
External factors are
disrupting the
enclosure airflow
patterns.
See “Location Requirements” Chapter 2,
“Certifying the Enclosure” Chapter 3, and
“Safety Precautions” Chapter 4 sections of
this manual. (Ref. Appendix D).
Enclosure has
improper downflow
velocity.
Have enclosure certified and check
exhaust system. Check HEPA filters for
loading. Adjust enclosure speed control.
Enclosure should have an average
downflow velocity of 30-65 fpm.
Blower won’t
operate.
Unit not plugged
into outlet.
Plug the enclosure into appropriate
electrical service.
- Purifier Filtered PCR Enclosures 3970422 Purifier Filtered PCR Enclosures 3970405 Purifier Filtered PCR Enclosures 3970402 Purifier Filtered PCR Enclosures 3970325 Purifier Filtered PCR Enclosures 3970322 Purifier Filtered PCR Enclosures 3970305 Purifier Filtered PCR Enclosures 3970302 Purifier Filtered PCR Enclosures 3970225 Purifier Filtered PCR Enclosures 3970222 Purifier Filtered PCR Enclosures 3970205 Purifier Filtered PCR Enclosures 3970202