Verifying hepa filter integrity – Labconco Purifier Filtered PCR Enclosures 3970425 User Manual
Page 15

Chapter 3: Getting Started
Product Service 1-800-522-7658
11
Verifying HEPA Filter Integrity
The HEPA filter is shipped installed with the gasket on the downstream side. The
HEPA filter is leak checked at Labconco. A second leak check is recommended
before using the enclosure and at least annually thereafter. Consult your Safety
Officer and Chapter 6 for the HEPA Filter Leak Test. See Figure 4-2 for HEPA
filter location, HEPA filter gasket, and filter clamp bolts. All seams downstream of
the HEPA filter are jacketed by positive pressure. This intrinsically safe design
ensures Class 5 conditions and a particulate free environment for PCR
experiments.
HEPA Filter
Appropriate Use
Filter Size
Part No.
HEPA filters are high-efficiency particulate
air filters having a particulate removable
efficiency of 99.99% for particles with a
diameter of 0.3 micron.
2'
3707900
3'
3707901
4'
3707902
Connecting the Electrical Supply Source to the
Filtered PCR Enclosure
115V Models
Simply connect the 115V power cord supplied to the IEC electrical supply plug on
the back of the enclosure. If using at 50 Hz operations, blower performance
maximum airflow will be reduced by 17%.
230V Models
The 230V is shipped without a plug. Install the appropriate plug for your electrical
specifications per local codes.
Do not use any detachable power cord that is not adequately rated
for the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.
!
- Purifier Filtered PCR Enclosures 3970422 Purifier Filtered PCR Enclosures 3970405 Purifier Filtered PCR Enclosures 3970402 Purifier Filtered PCR Enclosures 3970325 Purifier Filtered PCR Enclosures 3970322 Purifier Filtered PCR Enclosures 3970305 Purifier Filtered PCR Enclosures 3970302 Purifier Filtered PCR Enclosures 3970225 Purifier Filtered PCR Enclosures 3970222 Purifier Filtered PCR Enclosures 3970205 Purifier Filtered PCR Enclosures 3970202