Chapter 6: Maintaining Your Filtered PCR Enclosure
Product Service 1-800-522-7658
36
36
Speed Control Replacement
1. Remove the bracket that the speed control is attached to. See Figures 6-1, and
6-2.
2. Remove the two screws holding the speed control using a Phillips screwdriver.
Refer to Appendix A for Replacement Parts Diagram.
3. Disconnect all wires leading to the speed control. Connect wires on new speed
control in the same position as the old speed control.
4. Reassemble to the system in the same position and with the same screws that
were removed earlier.
This manual is related to the following products:
- Purifier Filtered PCR Enclosures 3970422 Purifier Filtered PCR Enclosures 3970405 Purifier Filtered PCR Enclosures 3970402 Purifier Filtered PCR Enclosures 3970325 Purifier Filtered PCR Enclosures 3970322 Purifier Filtered PCR Enclosures 3970305 Purifier Filtered PCR Enclosures 3970302 Purifier Filtered PCR Enclosures 3970225 Purifier Filtered PCR Enclosures 3970222 Purifier Filtered PCR Enclosures 3970205 Purifier Filtered PCR Enclosures 3970202