TROUBLESHOOTING

30

PROBLEM CAUSES

CORRECTIVE

ACTION

Water is noisy when
flowing through the
unit

Excessive water flow rate Reduce the water flow rate

slightly

Steam drains
erratically; steam
valve is noisy during
standby

Drain hose is kinked or
elevated above the drain
pipe

Drain hose must flow
unrestricted to a drain below the
drain pipe

Boiling flask breaks

Flask has gone dry

Do not let flask go dry

Water added when boiler
is dry or low on water and
hot

Do not add water to the boiler
when it is hot and dry, or low on
water

Recoveries are low

Viton seal is leaking

Clean/replace the Viton seal

Receiving tube is not
completely submerged

The receiving tube must be
completely submerged in the
receiving vessel

Not collecting enough
distillate

Increase distillation time

Recoveries are too
high

Excessive NH

3

in feed

water

Connect to a different water
source; install Optional Water
Inlet Kit #65442 and feed
purified water to the boiler