SAFETY PRECAUTIONS

15

The disposal of substances used in connection with this equipment may be governed by various
federal, state or local regulations. All users of this equipment are urged to become familiar with
any regulations that apply in the user’s area concerning the dumping of waste materials in or upon
water, land or air and to comply with such regulations.

The RapidStill II should not be operated in damp conditions or those suspect of an explosion and
should be situated to allow adequate room and ventilation during its normal usage.

The RapidStill II employs the following: electrical devices, heating element, hot water, steam,
pressurized water and steam lines. Care should be taken when installing, operating, or servicing
the unit.

Ensure that the RapidStill II is connected to electrical service in accordance to local and national
electrical codes. Failure to do so may create a fire or electrical hazard.

The RapidStill II is designed to be connected to a cold, potable water supply purified by municipal
water treatment facilities. The installation of the optional water inlet kit allows for the use of
purified water to supply the boiler. Never connect the RapidStill II to non-potable, unprocessed
surface or well water systems. The use of inadequate water supplies may damage the unit.

Always wear appropriate eye protection, protective gloves and clothing when using the RapidStill
II.

Never operate the RapidStill II without the Safety Screen and Safety Door in their proper position.

Never turn on the boiler heater when the boiler flask is dry or low on water. Do not let the boiler
run dry. Do not attempt to refill the boiler when it is dry or low on water and hot. Failure to
ensure proper water level in the boiler, or attempting to fill it when hot may result in the breakage
of the boiler flask and/or damage to the heating element.

At then end of the day, rinse the caustic delivery system with purified water. Remove any
accumulated spillage or deposits from the surfaces of the unit to prevent corrosion or damage.