TROUBLESHOOTING

29

PROBLEM CAUSES

CORRECTIVE

ACTION

Unit will not operate
at all

Tripped circuit breaker

Check facility’s circuit breaker

Blown fuse

Replace fuse

Power cord not plugged
in

Plug in cord

Boiler doesn’t fill

‘Boiler Heater’ switch is
on

Turn off boiler heater before
filling

Cooling water off or flow
rate insufficient

Check flow rate of condenser
cooling water

Water fill valve plugged
or defective

Clean inlet screen of filter;
replace valve

‘Boiler Water’ switch
defective

Replace switch

Heating element
doesn’t work

Heating element is
defective

Replace heating element

‘Boiler Heater’ switch is
defective

Replace switch

NaOH pump doesn’t
run

‘Boiler Heater’ switch is
off

The boiler heater must be on for
the NaOH pump to operate

Distillation Timer is set in
the timing mode

The Distillation timer must be
set to ‘0’ for the NaOH pump to
work

NaOH pump is defective

Replace the NaOH pump

NaOH pump runs,
but doesn’t pump

NaOH tube is kinked or
restricted

Check all lines for kinks or
blockage

Peristaltic pump tube
damaged

Replace peristaltic pump tube

Peristaltic pump is
damaged

Replace peristaltic pump