INSTALLATION

10

Location

The RapidStill II should be located on a level work surface, adjacent to appropriate electrical,
water and drain connections.

Electrical Connection

The electrical outlet should be rated for 115 VAC, 60 Hz, 15 Amps, and dedicated to the
RapidStill II.

Water Connection

The feed water should be municipal drinking water grade and of sufficient quality to prevent or
minimize scaling of the boiler. The feed water should have a minimum supply volume of 1 gallon
(3.8 liters) per minute, at 10 PSI.

Using the supplied braided water line, connect the tube to the water inlet fitting on the rear of the
RapidStill II as shown in Figure 4. To connect or disconnect the fittings, follow the instructions in
Figure 5, below. If your feed water promotes excessive scaling of the boiler, or has high NH

3

levels, you may wish to install the Optional Water Inlet Kit, part #65442, described on page 12.